NBS Silicon Powder
Silicon can be used to calibrate WAXS detectors (e.g. INEL, MSGC). The prominent peaks are listed below the scattering curve.

| H K L | d-spacing/Å | Angle/° |
| 111 | 3.1355 | 9.14628 |
| 220 | 1.9201 | 14.96249 |
| 311 | 1.6374 | 17.56393 |
| 222 | 1.5677 | 18.35156 |
| 400 | 1.3577 | 21.22116 |
| 331 | 1.2459 | 23.15043 |
| 422 | 1.1085 | 26.06643 |
| 333 | 1.0451 | 27.67815 |
| 511 | 1.0451 | 27.67815 |
| 440 | 0.9600 | 30.18796 |
| 531 | 0.9179 | 31.60668 |
| 442 | 0.9051 | 32.0635 |
| 620 | 0.8586 | 33.85153 |
| 533 | 0.8281 | 35.13871 |
| 622 | 0.8187 | 35.55841 |
| 444 | 0.7838 | 37.19614 |
| 711 | 0.7604 | 38.38497 |
| 551 | 0.7604 | 38.38497 |
| 642 | 0.7257 | 40.30023 |
For more information on Si standards one is referred to the website of NIST.
(Note: the diffraction angles refer to experiments with lambda = 0.5 Å. Where NIST has used Cu Kα radiation.)