| Scalar Attributes |
| Attribute name |
Data Type |
R/W Type |
Expert |
| integrationConstant: The integration constant sets the duration of the digital averaging carried out on acquisitions to 2kHz. The possible values are: 0;1;2;5;10;20;50;100;200;500;1000. Do not confuse the integration constant with the sample period which sets the time interval separating two successive measurements in dynamic acquisition mode (TP,TK). |
DEV_DOUBLE | READ_WRITE | No |
| angleY: is the Y-angle measurement. This value is the average over nb samples. It corresponds to a rotation around the y-axis. Axis references are defined in the user's manual. The range of the magnitude depend on the head used. |
DEV_DOUBLE | READ | No |
| angleZ: is Z-angle measurement. This value is the average over nb samples. It corresponds to a rotation around the y-axis. Axis references are defined in the user's manual. The range of the magnitude depend on the head used. |
DEV_DOUBLE | READ | No |
| countingCompleted: gives an indication about the process of acquiring. 2 values are possible 0 (running) and 1(standbye) The device scanservers needs it. |
DEV_SHORT | READ | No |
| angleYStdDeviation: Provides the mean standard deviation for Y angle |
DEV_DOUBLE | READ | No |
| angleZStdDeviation: Provides the mean standard deviation for Z angle |
DEV_DOUBLE | READ | No |
| numberSamples: Number of samples for each measurement |
DEV_LONG | READ_WRITE | No |
| samplePeriod: Sets the sampling period to X ms |
DEV_DOUBLE | READ_WRITE | No |
| integrationTime: Not used (only for Compatibility with ScanServer) |
DEV_DOUBLE | WRITE | No |