TANGO
Device Server



LDS1000
Device Attributes Description

LDS1000 Class

Revision: release_1_0_4 - Author: vince_soleil

Scalar Attributes
Attribute name
Data Type
R/W Type
Expert
integrationConstant: The integration constant sets the duration of the digital averaging carried out on acquisitions to 2kHz. The possible values are: 0;1;2;5;10;20;50;100;200;500;1000. Do not confuse the integration constant with the sample period which sets the time interval separating two successive measurements in dynamic acquisition mode (TP,TK).
DEV_DOUBLE
READ_WRITE
No
angleY: is the Y-angle measurement. This value is the average over nb samples. It corresponds to a rotation around the y-axis. Axis references are defined in the user's manual. The range of the magnitude depend on the head used.
DEV_DOUBLE
READ
No
angleZ: is Z-angle measurement. This value is the average over nb samples. It corresponds to a rotation around the y-axis. Axis references are defined in the user's manual. The range of the magnitude depend on the head used.
DEV_DOUBLE
READ
No
countingCompleted: gives an indication about the process of acquiring. 2 values are possible 0 (running) and 1(standbye) The device scanservers needs it.
DEV_SHORT
READ
No
angleYStdDeviation: Provides the mean standard deviation for Y angle
DEV_DOUBLE
READ
No
angleZStdDeviation: Provides the mean standard deviation for Z angle
DEV_DOUBLE
READ
No
numberSamples: Number of samples for each measurement
DEV_LONG
READ_WRITE
No
samplePeriod: Sets the sampling period to X ms
DEV_DOUBLE
READ_WRITE
No
integrationTime: Not used (only for Compatibility with ScanServer)
DEV_DOUBLE
WRITE